Show Nav →
Please include the following statement when referencing the CPTAC Assay Portal
We would like to acknowledge the National Cancer Institute’s Clinical Proteomic Tumor Analysis Consortium (CPTAC) Assay Portal (assays.cancer.gov) for developing assays and establishing criteria for the assays described in this publication.

Overview Data source: UniProt

Official Gene Symbol Other Aliases
CCT3 CCTG, TRIC5
Sequence Length (AA) Molecular Weight (Da)
545 60534
Protein Name
T-complex protein 1 subunit gamma
Sources
UniProt
PhosphoSitePlus ®
GeneCards
Human Protein Atlas

Protein Sequence hover to view complete sequence

10 20 30 40 50
MMGHRPVLVL SQNTKRESGR KVQSGNINAA KTIADIIRTC LGPKSMMKML
60 70 80 90 100
LDPMGGIVMT NDGNAILREI QVQHPAAKSM IEISRTQDEE VGDGTTSVII
110 120 130 140 150
LAGEMLSVAE HFLEQQMHPT VVISAYRKAL DDMISTLKKI SIPVDISDSD
160 170 180 190 200
MMLNIINSSI TTKAISRWSS LACNIALDAV KMVQFEENGR KEIDIKKYAR
210 220 230 240 250
VEKIPGGIIE DSCVLRGVMI NKDVTHPRMR RYIKNPRIVL LDSSLEYKKG
260 270 280 290 300
ESQTDIEITR EEDFTRILQM EEEYIQQLCE DIIQLKPDVV ITEKGISDLA
310 320 330 340 350
QHYLMRANIT AIRRVRKTDN NRIARACGAR IVSRPEELRE DDVGTGAGLL
360 370 380 390 400
EIKKIGDEYF TFITDCKDPK ACTILLRGAS KEILSEVERN LQDAMQVCRN
410 420 430 440 450
VLLDPQLVPG GGASEMAVAH ALTEKSKAMT GVEQWPYRAV AQALEVIPRT
460 470 480 490 500
LIQNCGASTI RLLTSLRAKH TQENCETWGV NGETGTLVDM KELGIWEPLA
510 520 530 540 545
VKLQTYKTAV ETAVLLLRID DIVSGHKKKG DDQSRQGGAP DAGQE

Data source: UniProt


Position of Targeted Peptide Analytes Relative to SNPs, Isoforms, and PTMs

Uniprot Database Entry PhosphoSitePlus ®

Click a point on a node
to view detailed assay information below
All other points link out to UniProt



Phosphorylation Acetylation Ubiquitylation Other

loading

Loader

Assay Details for non-CPTAC-6084 Collapse assay details

Data source: Panorama

Official Gene Symbol
CCT3
Peptide Sequence
ELGIWEPLAVK
Modification Type
unmodified
Protein - Site of Modification
N/A
Peptide - Site of Modification
N/A
Peptide Start
492
Peptide End
502
CPTAC ID
non-CPTAC-6084
Peptide Molecular Mass
1,253.7020
Species
Homo sapiens (Human)
Assay Type
Direct MRM-MS
Matrix
Bovine Serum Albumin
Submitting Laboratory
McGill University
Submitting Lab PI
Christoph H. Borchers, Gerald Batist

Publication

View Details (opens in a new window)

mTORC1-driven protein translation correlates with clinical benefit of capivasertib within a genetically preselected cohort of PIK3CA-altered tumours. Constance A Sobsey, Bjoern C Froehlich, Georgia Mitsa, Sahar Ibrahim, Robert Popp, Rene P Zahedi, Elza C de Bruin, Christoph H Borchers, Gerald Batist


Assay Parameters Collapse assay parameters

Data source: Panorama

Instrument
Agilent 6495B QQQ-MS
Internal Standard
Standard Isotope labeled internal standard peptides with 13Cx, 15Ny (R+10 or K+8)
Peptide Standard Purity
>95%
Peptide Standard Label Type
13C and 15N at C-terminus K
LC
Infinity 1290
Column Packing
commercial (Agilent ZORBAX Eclipse Plus C18)
Column Dimensions
4.6 mm × 150 mm, (5 µm)
Flow Rate
0.4 mL/min

Chromatograms

Data source: Panorama


Response Curves

Data source: Panorama

Retrieving Data

Loader

Repeatability

Data source: Panorama

  Average intra-assay CV
(within day CV)
Average inter-assay CV
(between day CV)
Total CV
equation
n=
Fragment ion / Transition Low Med High Low Med High Low Med High Low Med High
b3 (1+) 8.9 5.1 8.1 8.9 5.6 8.5 12.6 7.6 11.7 16 16 17
y5 (1+) 4.5 4.2 5.7 5.7 5.4 6.5 7.3 6.8 8.6 16 16 17
y6 (1+) 11 4.2 4.9 9.1 5 4.9 14.3 6.5 6.9 16 16 17
y7 (1+) 10.3 5.8 4.1 7.3 5.6 3.3 12.6 8.1 5.3 16 16 17
y9 (1+) 5.6 5.2 4.6 4.6 5.3 4.5 7.2 7.4 6.4 16 16 17
sum 4.9 2.2 2.3 2.6 2.3 1.3 5.5 3.2 2.6 16 16 17


Stability

Data source: Panorama

Fragment ion / Transition control_intra_CV actual_temp_intra_CV frozen_intra_CV FTx1_intra_CV FTx2_intra_CV
b3 (1+) 17.6 4.2 3.9 4.3 8.8
y5 (1+) 7.4 5.4 3.3 3 4.5
y6 (1+) 9.5 6.5 4.4 7.7 2.6
y7 (1+) 8.6 4.1 3.8 6.1 4.2
y9 (1+) 9.3 7 3.8 2.4 2
sum 2.9 1.2 1.4 0.9 2.8
Fragment ion / Transition all_intra_CV all_inter_CV
b3 (1+) 7.1 8.4
y5 (1+) 4.8 5.8
y6 (1+) 6.2 6.2
y7 (1+) 5.1 6.1
y9 (1+) 5.2 6.4
sum 1.7 2.4

Endogenous

Data source: Panorama

Fragment ion / Transition intra_CV inter_CV total_CV
b3 (1+) 15.8 26 30.4
y5 (1+) 8.1 23.4 24.8
y6 (1+) 6.7 19.9 21
y7 (1+) 18.5 18.2 26
y9 (1+) 3 29.9 30.1
sum 5.3 21.4 22

Additional Resources and Comments


Assay Details for CPTAC-475 Collapse assay details

Data source: Panorama

Official Gene Symbol
CCT3
Peptide Sequence
ELGIWEPLAVK
Modification Type
unmodified
Protein - Site of Modification
N/A
Peptide - Site of Modification
N/A
Peptide Start
492
Peptide End
502
CPTAC ID
CPTAC-475
Peptide Molecular Mass
1,253.7020
Species
Homo sapiens (Human)
Assay Type
Direct MRM-MS
Matrix
cell line lysate pool
Submitting Laboratory
Seoul National University / Korea Institute of Science and Technology and FHCRC
Submitting Lab PI
Youngsoo Kim

Publication

View Details (opens in a new window)

Demonstrating the feasibility of large-scale development of standardized assays to quantify human proteins. Kennedy JJ, Abbatiello SE, Kim K, Yan P, Whiteaker JR, Lin C, Kim JS, Zhang Y, Wang X, Ivey RG, Zhao L, Min H, Lee Y, Yu MH, Yang EG, Lee C, Wang P, Rodriguez H, Kim Y, Carr SA, Paulovich AG. Nat Methods. 2014 Feb;11(2):149-55. doi: 10.1038/nmeth.2763. Epub 2013 Dec 8. PMID: 24317253


Response curves acquired at Seoul National University / Korea Institute of Science and Technology. Repeatability data acquired at Fred Hutchinson Cancer Research Center.


Response curves acquired at Seoul National University / Korea Institute of Science and Technology. Repeatability data acquired at Fred Hutchinson Cancer Research Center.


Assay Parameters Collapse assay parameters

Data source: Panorama

Instrument
5500 QTRAP (ABSCIEX)
Internal Standard
peptide
Peptide Standard Purity
>95%
Peptide Standard Label Type
13C and 15N at C-terminus K
LC
nanoLC-Ultra 2D, cHiPLC-nanoflex (Eksigent)
Column Packing
ChromXP C18-CL, 3 um, 120A
Column Dimensions
150 x 0.075 mm
Flow Rate
500 nL / min

Chromatograms

Data source: Panorama


Response Curves

Data source: Panorama

Retrieving Data

Loader

Repeatability

Data source: Panorama

  Average intra-assay CV
(within day CV)
Average inter-assay CV
(between day CV)
Total CV
equation
n=
Fragment ion / Transition Low Med High Low Med High Low Med High Low Med High
y9 (1+) 13.7 7.7 5.7 21.4 7.8 6.4 25.4 11 8.6 15 25 15
y7 (1+) 7.3 7.4 5.8 17.5 8.1 6.4 19 11 8.6 15 25 15
y6 (1+) 15.4 8.5 4.6 16.3 8.6 7.1 22.4 12.1 8.5 15 25 15
sum 8.4 6.7 5.2 16.7 7.2 6.1 18.7 9.8 8 15 25 15


Additional Resources and Comments


Assay Details for CPTAC-474 Collapse assay details

Data source: Panorama

Official Gene Symbol
CCT3
Peptide Modified Sequence
IPGGIIEDSC[+57.0]VLR
Modification Type
Carbamidomethyl Cysteine
Protein - Site of Modification
213
Peptide - Site of Modification
10
Peptide Start
204
Peptide End
216
CPTAC ID
CPTAC-474
Peptide Molecular Mass
1,427.7443
Species
Homo sapiens (Human)
Assay Type
Direct MRM-MS
Matrix
cell line lysate pool
Submitting Laboratory
Seoul National University / Korea Institute of Science and Technology and FHCRC
Submitting Lab PI
Youngsoo Kim

Publication

View Details (opens in a new window)

Demonstrating the feasibility of large-scale development of standardized assays to quantify human proteins. Kennedy JJ, Abbatiello SE, Kim K, Yan P, Whiteaker JR, Lin C, Kim JS, Zhang Y, Wang X, Ivey RG, Zhao L, Min H, Lee Y, Yu MH, Yang EG, Lee C, Wang P, Rodriguez H, Kim Y, Carr SA, Paulovich AG. Nat Methods. 2014 Feb;11(2):149-55. doi: 10.1038/nmeth.2763. Epub 2013 Dec 8. PMID: 24317253


Response curves acquired at Seoul National University / Korea Institute of Science and Technology. Repeatability data acquired at Fred Hutchinson Cancer Research Center.


Response curves acquired at Seoul National University / Korea Institute of Science and Technology. Repeatability data acquired at Fred Hutchinson Cancer Research Center.


Assay Parameters Collapse assay parameters

Data source: Panorama

Instrument
5500 QTRAP (ABSCIEX)
Internal Standard
peptide
Peptide Standard Purity
>95%
Peptide Standard Label Type
13C,15N
LC
nanoLC-Ultra 2D, cHiPLC-nanoflex (Eksigent)
Column Packing
ChromXP C18-CL, 3 um, 120A
Column Dimensions
150 x 0.075 mm
Flow Rate
500 nL / min

Chromatograms

Data source: Panorama


Response Curves

Data source: Panorama

Retrieving Data

Loader

Repeatability

Data source: Panorama

  Average intra-assay CV
(within day CV)
Average inter-assay CV
(between day CV)
Total CV
equation
n=
Fragment ion / Transition Low Med High Low Med High Low Med High Low Med High
y11 (1+) 10.8 8.9 4.2 15.3 11.2 5.8 18.7 14.3 7.2 15 25 15
y8 (1+) 9.1 9.5 4.9 13.3 10.3 7.5 16.1 14 9 15 25 15
y7 (1+) 8.7 8.8 3.1 13 10.3 7.4 15.6 13.5 8 15 25 15
sum 8.1 8.8 3.5 13.4 10.3 6.2 15.7 13.5 7.1 15 25 15


Additional Resources and Comments