Show Nav →
Please include the following statement when referencing the CPTAC Assay Portal
We would like to acknowledge the National Cancer Institute’s Clinical Proteomic Tumor Analysis Consortium (CPTAC) Assay Portal (assays.cancer.gov) for developing assays and establishing criteria for the assays described in this publication.

Overview Data source: UniProt

Official Gene Symbol Other Aliases
Tfpi2 N/A
Sequence Length (AA) Molecular Weight (Da)
230 26137
Protein Name
Tissue factor pathway inhibitor 2
Sources
UniProt
PhosphoSitePlus ®
GeneCards
Human Protein Atlas

Protein Sequence hover to view complete sequence

10 20 30 40 50
MDPAMPLQLW NLPLLLVGSV LGLTSVSAQG NNLEICLLPL DAGPCQALIP
60 70 80 90 100
KFYYDRDQQK CRRFNYGGCL GNANNFHSRD LCQQTCGSIE KVPPVCRSEL
110 120 130 140 150
KTYPCDKPNI RFFFNLNTMT CEPLRPGLCS RTINVFSEEA TCKGLCEPRK
160 170 180 190 200
HIPSFCSSPK DEGLCSANVT RFYFNSRNKT CETFTYTGCG GNENNFYYLD
210 220 230
ACHRACVKGW KKPKRWKIGD FLPRFWKHLS

Data source: UniProt


Position of Targeted Peptide Analytes Relative to SNPs, Isoforms, and PTMs

Uniprot Database Entry PhosphoSitePlus ®

Click a point on a node
to view detailed assay information below
All other points link out to UniProt



Phosphorylation Acetylation Ubiquitylation Other

loading

Loader

Assay Details for non-CPTAC-3502 Collapse assay details

Data source: Panorama

Official Gene Symbol
Tfpi2
Peptide Sequence
IGDFLPR
Modification Type
unmodified
Protein - Site of Modification
N/A
Peptide - Site of Modification
N/A
Peptide Start
218
Peptide End
224
CPTAC ID
non-CPTAC-3502
Peptide Molecular Mass
816.4494
Species
Mus musculus (Mouse)
Assay Type
Direct MRM
Matrix
Plasma
Submitting Laboratory
UVic-Genome BC Proteomics Centre
Submitting Lab PI
Christoph Borchers

Assay Parameters Collapse assay parameters

Data source: Panorama

Instrument
Agilent 6490/6495 QQQ
Internal Standard
synthetic peptide
Peptide Standard Purity
>95%
Peptide Standard Label Type
13C and 15N at C-terminus R
LC
1290 LC (Agilent)
Column Packing
Zorbax Eclipse Plus C18, 1.8 µm
Column Dimensions
2.1 x 150 mm
Flow Rate
400 µL/min

Chromatograms

Data source: Panorama


Response Curves

Data source: Panorama

Retrieving Data

Loader

Repeatability

Data source: Panorama

  Average intra-assay CV
(within day CV)
Average inter-assay CV
(between day CV)
Total CV
equation
n=
Fragment ion / Transition Low Med High Low Med High Low Med High Low Med High
y2 (1+) 9.1 4.4 4.6 12.6 6.2 6.3 15.5 7.6 7.8 15 15 15
y3 (1+) 25.2 9.7 6.2 28.6 7.9 9.1 38.1 12.5 11 15 15 15
y4 (1+) 4.8 3.9 2.7 6.6 4.5 4.5 8.2 6 5.2 15 15 15
y5 (1+) 4.5 2.9 3.2 8.9 4.9 3.9 10 5.7 5 15 15 15
y6 (1+) 6.1 2.6 2.8 9 4.4 5.2 10.9 5.1 5.9 15 15 15
sum 3.9 1.5 1.5 6.8 4.1 4.1 7.8 4.4 4.4 15 15 15


Additional Resources and Comments