Overview Data source: UniProt

Official Gene Symbol Other Aliases
Rpl14 N/A
Sequence Length (AA) Molecular Weight (Da)
217 23564
Protein Name
60S ribosomal protein L14
Sources
UniProt
PhosphoSitePlus ®
GeneCards
Human Protein Atlas

Protein Sequence hover to view complete sequence

10 20 30 40 50
MVFRRYVEVG RVAYISFGPH AGKLVAIVDV IDQNRALVDG PCTRVRRQAM
60 70 80 90 100
PFKCMQLTDF ILKFPHSARQ KYVRKAWEKA DINTKWAATR WAKKIDARER
110 120 130 140 150
KAKMTDFDRF KVMKAKKMRN RIIKTEVKKL QRAAILKASP KKAAVAKAAI
160 170 180 190 200
AAAAAAAAAK AKVPAKKATG PGKKAAGQKA PAQKAAGQKA APPAKGQKGQ
210 217
KTPAQKAPAP KAAGKKA

Data source: UniProt


Position of Targeted Peptide Analytes Relative to SNPs, Isoforms, and PTMs

Uniprot Database Entry PhosphoSitePlus ®

Click a point on a node
to view detailed assay information below
All other points link out to UniProt



Phosphorylation Acetylation Ubiquitylation Other

loading

Loader

Assay Details for non-CPTAC-3667 Collapse assay details

Data source: Panorama

Official Gene Symbol
Rpl14
Peptide Sequence
LVAIVDVIDQNR
Modification Type
unmodified
Protein - Site of Modification
N/A
Peptide - Site of Modification
N/A
Peptide Start
24
Peptide End
35
CPTAC ID
non-CPTAC-3667
Peptide Molecular Mass
1,353.7616
Species
Mus musculus (Mouse)
Assay Type
Direct MRM
Matrix
Plasma
Submitting Laboratory
UVic-Genome BC Proteomics Centre
Submitting Lab PI
Christoph Borchers

Assay Parameters Collapse assay parameters

Data source: Panorama

Instrument
Agilent 6490/6495 QQQ
Internal Standard
synthetic peptide
Peptide Standard Purity
>95%
Peptide Standard Label Type
13C and 15N at C-terminus R
LC
1290 LC (Agilent)
Column Packing
Zorbax Eclipse Plus C18, 1.8 µm
Column Dimensions
2.1 x 150 mm
Flow Rate
400 µL/min

Chromatograms

Data source: Panorama


Response Curves

Data source: Panorama

Retrieving Data

Loader

Repeatability

Data source: Panorama

  Average intra-assay CV
(within day CV)
Average inter-assay CV
(between day CV)
Total CV
equation
n=
Fragment ion / Transition Low Med High Low Med High Low Med High Low Med High
b4 (1+) 13.4 7.2 4.2 14.1 11.6 10.4 19.5 13.7 11.2 15 15 15
b3 (1+) 7.9 3.3 1.6 10.5 13.1 10.3 13.1 13.5 10.4 15 15 15
y3 (1+) 26.9 8.3 5.1 29.4 18.4 10 39.8 20.2 11.2 15 15 15
y7 (1+) 9.2 4.1 4.3 7.9 15.7 8.5 12.1 16.2 9.5 15 15 15
y8 (1+) 9.6 4.2 3.4 11.9 11.2 9.9 15.3 12 10.5 15 15 15
sum 5.8 2.5 1.9 6.7 12.4 9.6 8.9 12.6 9.8 15 15 15


Additional Resources and Comments