Show Nav →
Please include the following statement when referencing the CPTAC Assay Portal
We would like to acknowledge the National Cancer Institute’s Clinical Proteomic Tumor Analysis Consortium (CPTAC) Assay Portal (assays.cancer.gov) for developing assays and establishing criteria for the assays described in this publication.

Overview Data source: UniProt

Official Gene Symbol Other Aliases
CD70 CD27L, CD27LG, TNFSF7
Sequence Length (AA) Molecular Weight (Da)
193 21118
Protein Name
CD70 antigen
Sources
UniProt
PhosphoSitePlus ®
GeneCards
Human Protein Atlas

Protein Sequence hover to view complete sequence

10 20 30 40 50
MPEEGSGCSV RRRPYGCVLR AALVPLVAGL VICLVVCIQR FAQAQQQLPL
60 70 80 90 100
ESLGWDVAEL QLNHTGPQQD PRLYWQGGPA LGRSFLHGPE LDKGQLRIHR
110 120 130 140 150
DGIYMVHIQV TLAICSSTTA SRHHPTTLAV GICSPASRSI SLLRLSFHQG
160 170 180 190 193
CTIASQRLTP LARGDTLCTN LTGTLLPSRN TDETFFGVQW VRP

Data source: UniProt


Position of Targeted Peptide Analytes Relative to SNPs, Isoforms, and PTMs

Uniprot Database Entry PhosphoSitePlus ®

Click a point on a node
to view detailed assay information below
All other points link out to UniProt



Phosphorylation Acetylation Ubiquitylation Other

loading

Loader

Assay Details for CPTAC-5959 Collapse assay details

Data source: Panorama

Official Gene Symbol
CD70
Peptide Sequence
LYWQGGPALGR
Modification Type
unmodified
Protein - Site of Modification
N/A
Peptide - Site of Modification
4
Peptide Start
73
Peptide End
83
CPTAC ID
CPTAC-5959
Peptide Molecular Mass
1,216.6353
Species
Homo sapiens (Human)
Assay Type
Enrichment MRM
Enrichment Method
peptide immunoaffinity
Matrix
FFPE tumor tissue lysate pool
Submitting Laboratory
Fred Hutchinson Cancer Research Center
Submitting Lab PI
Amanda Paulovich

Publication

View Details (opens in a new window)

Targeted Mass Spectrometry Enables Multiplexed Quantification of Immunomodulatory Proteins in Clinical Biospecimens.Whiteaker JR, Lundeen RA, Zhao L, Schoenherr RM, Burian A, Huang D, Voytovich U, Wang T, Kennedy JJ, Ivey RG, Lin C, Murillo OD, Lorentzen TD, Thiagarajan M, Colantonio S, Caceres TW, Roberts RR, Knotts JG, Reading JJ, Kaczmarczyk JA, Richardson CW, Garcia-Buntley SS, Bocik W, Hewitt SM, Murray KE, Do N, Brophy M, Wilz SW, Yu H, Ajjarapu S, Boja E, Hiltke T, Rodriguez H, Paulovich AG.Front Immunol. 2021 Nov 11;12:765898. doi: 10.3389/fimmu.2021.765898. eCollection 2021.PMID: 34858420


Assay Parameters Collapse assay parameters

Data source: Panorama

Instrument
QTRAP 5500
Internal Standard
peptide
Peptide Standard Purity
>95%
Peptide Standard Label Type
13C and 15N at C-terminus R
LC
Eksigent nanoLC 425
Column Packing
3 µm ChromXP C18EP, 120 Å
Column Dimensions
150 x 0.3 mm
Flow Rate
0.3 µL/min

Chromatograms

Data source: Panorama


Response Curves

Data source: Panorama

Retrieving Data

Loader

Repeatability

Data source: Panorama

  Average intra-assay CV
(within day CV)
Average inter-assay CV
(between day CV)
Total CV
equation
n=
Fragment ion / Transition Low Med High Low Med High Low Med High Low Med High
y9 (1+) 9.4 6.8 4.2 11.5 8.6 14.9 14.9 11 15.5 13 25 14
b2 (1+) 10.6 5.3 5.1 8.8 12.2 11.8 13.8 13.3 12.9 13 25 14
y9 (2+) 5.9 5.4 4.4 9.9 10.8 11.7 11.5 12.1 12.5 13 25 14
y8 (1+) 7 6 2.8 8.8 10.7 12.3 11.2 12.3 12.6 13 25 14
y7 (1+) 5.8 5.4 4.4 7.6 10.2 11.8 9.6 11.5 12.6 13 25 14
y6 (1+) 11.1 7.2 3.9 12.6 11.3 14.3 16.8 13.4 14.8 13 25 14
y2 (1+) 10.4 5.4 6.3 13.2 10.8 14.4 16.8 12.1 15.7 13 25 14
sum 6.4 4.6 2.5 8 10.1 12 10.2 11.1 12.3 13 25 14


Additional Resources and Comments